Veterinary Anaesthesia Machine with Ventilator
Features:
*precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate.
*one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
*Reliable pneumatic transmission system.
*Economic and proper configuration of technologies, optional as required by the Customer.
*Modularized design and upgrading available as required by the Customer.
Part of Anesthesia
Main body : High-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of application : adult, child
Gas source : O2, N2O: 0.27-0.55MPa
Flow meter : O2: 0.1-1.0L/min 1.1-10L/min
Flow rate of rapid oxygen supply : 25-75L/min
Low oxygen pressure alarm sound alarm when the oxygen pressure < 0.2MPa
Vaporizer : It has the function of automatic compensation based on pressure,temperature,and flow rate. The regulation range of evaporator concentration is 0-5 vol%.
Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by customer.
Respiratory circuit working mode: all-close, semi-close, semi-open releasing pressure: 0.5-7 kPa
Respiratory bellows :bellows for adults, bellows for children tidal volume range:0-1500 ml
Part of Ventilator
Display mode : High-definition 4.8" LCD screen display
Ventilation mode : IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate)
Ventilation parameters
Tidal volume : 50-1500ml
Rate : 4-80 bpm
I:E 2:1-1:4
Pressure range :1.0-6.0 kPa
Parameters for ventilation monitoring : Tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform
Security Alarm System
Airway pressure alarm upper limit setting range 1.0-6.0 kPa
low limit setting range 0.4-2.0 kPa
TV alarm :upper limit setting range 0.5-2.0 , Low limit setting range 0.02-1.0L
Sustained high-pressure alarm : it will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there is no tidal volume input in 15s
Power alarm
Power : AC 220V±10% 50Hz±1Hz